Important Notices

URGENT MEDICAL DEVICE CORRECTION: Accu-Chek Aviva Plus test strips

Important information on Accu-Chek® Aviva Plus test strips with lot numbers 497864 and 498315: These affected test strip lots show an increased potential for the inability to dose, resulting in the temporary unavailability of blood glucose measurement. Discontinue use of strips from these lot numbers and follow all instructions in the link below.

19-003 (December 2019)

URGENT MEDICAL DEVICE CORRECTION: Accu-Chek Aviva Plus meter

Important information on Accu-Chek® Aviva Plus blood glucose meters with serial numbers either between 45505000001 and 45506335297 or between 45520000000 and 45523284925 potentially having unexpected short battery life or not powering on.

19-002 (November 2019)

URGENT MEDICAL DEVICE CORRECTION: Accu-Chek Guide meter

Important information on Accu-Chek® Guide blood glucose meters with serial numbers below 92311000001 potentially giving E-9 errors, unexpected low battery icons, having short battery life, or not powering on.

19-001 (October 2019)